• DocumentCode
    2451434
  • Title

    The electrophysiological signals recording

  • Author

    Golescu, N. ; Ciachir, G. ; Babarada, F. ; Arhip, J.

  • Author_Institution
    Fac. of Electron. Telecommun. & Inf. Technol., Univ. Politeh. of Bucharest, Romania
  • Volume
    2
  • fYear
    2009
  • fDate
    12-14 Oct. 2009
  • Firstpage
    553
  • Lastpage
    556
  • Abstract
    The paper presents the main aspects of design and realization of a chain for electrophysiological signal recording, respectively collection, processing and acquisition. The dynamic range compressor circuit or the automatic gain control contains a recovery delay to minimize the distortions, thermal behaviour compensation and an adaptation stage for levelmeter. After compressor was added a clipper, which must catch all the transitions, that escapes out of the dynamic range compressor and a lowpass filter that cuts abruptly the high frequencies. The data vector acquisition is performing by strong internal resources microcontroller including ten bits A/D conversion port. The signal processing was designed in order to be implemented in a custom design IC.
  • Keywords
    analogue-digital conversion; bioelectric phenomena; low-pass filters; medical signal detection; medical signal processing; A/D conversion; automatic gain control; data vector acquisition; electrophysiological signals recording; lowpass filter; signal acquisition; signal collection; signal processing; thermal behaviour compensation; Circuits; Delay; Dynamic range; Filters; Frequency; Gain control; Microcontrollers; Process design; Signal design; Signal processing; bioelectronics; custom design;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Semiconductor Conference, 2009. CAS 2009. International
  • Conference_Location
    Sinaia
  • ISSN
    1545-827X
  • Print_ISBN
    978-1-4244-4413-7
  • Type

    conf

  • DOI
    10.1109/SMICND.2009.5336652
  • Filename
    5336652