• DocumentCode
    2452912
  • Title

    Absolute determination of loss angle

  • Author

    Eklund, G.

  • Author_Institution
    Swedish Nat. Testing & Res. Inst., Boras, Sweden
  • fYear
    1996
  • fDate
    17-21 June 1996
  • Firstpage
    390
  • Lastpage
    391
  • Abstract
    A guard ring capacitor has been constructed and is used for absolute determinations of the loss angle of capacitors at SP (Swedish National Testing and Research Institute). The accuracy is estimated to /spl plusmn/1 ppm (k=2). The main source of error is from the quadrature errors in the capacitance bridge.
  • Keywords
    bridge circuits; bridge instruments; capacitors; dielectric loss measurement; electron device testing; loss-angle measurement; measurement errors; 20 to 100 pF; Swedish National Testing and Research Institute; accuracy; capacitance bridge; capacitor loss angle; guard ring capacitor; loss angle measurement; quadrature errors; Bridge circuits; Capacitance; Capacitors; Dielectric losses; Electrodes; Equations; Loss measurement; Resistors; Surface resistance; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 1996 Conference on
  • Conference_Location
    Braunschweig, Germany
  • Print_ISBN
    0-7803-3376-4
  • Type

    conf

  • DOI
    10.1109/CPEM.1996.547129
  • Filename
    547129