Title
Absolute determination of loss angle
Author
Eklund, G.
Author_Institution
Swedish Nat. Testing & Res. Inst., Boras, Sweden
fYear
1996
fDate
17-21 June 1996
Firstpage
390
Lastpage
391
Abstract
A guard ring capacitor has been constructed and is used for absolute determinations of the loss angle of capacitors at SP (Swedish National Testing and Research Institute). The accuracy is estimated to /spl plusmn/1 ppm (k=2). The main source of error is from the quadrature errors in the capacitance bridge.
Keywords
bridge circuits; bridge instruments; capacitors; dielectric loss measurement; electron device testing; loss-angle measurement; measurement errors; 20 to 100 pF; Swedish National Testing and Research Institute; accuracy; capacitance bridge; capacitor loss angle; guard ring capacitor; loss angle measurement; quadrature errors; Bridge circuits; Capacitance; Capacitors; Dielectric losses; Electrodes; Equations; Loss measurement; Resistors; Surface resistance; Testing;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements Digest, 1996 Conference on
Conference_Location
Braunschweig, Germany
Print_ISBN
0-7803-3376-4
Type
conf
DOI
10.1109/CPEM.1996.547129