• DocumentCode
    2459341
  • Title

    An isothermal method for coaxial microcalorimeter

  • Author

    Chie-Mien Chi

  • Author_Institution
    Center for Meas. Stand., Ind. Technol. Res. Inst., Hsinchu, Taiwan
  • fYear
    1996
  • fDate
    17-21 June 1996
  • Firstpage
    454
  • Lastpage
    455
  • Abstract
    An isothermal method is presented to reduce the thermal time constant of microcalorimeter. The method is a comparatively more efficient process of determining the effective efficiency of thermistor mount than classical measurement method. Using this technique with APC-7 coaxial microcalorimeter has substantially reduced the measurement time by a factor of six.
  • Keywords
    calorimeters; measurement standards; microwave measurement; thermistors; APC-7; coaxial microcalorimeter; effective efficiency; isothermal method; measurement time; thermal time constant; thermistor mount; Automatic control; Bolometers; Coaxial components; Electromagnetic heating; Isothermal processes; Microwave theory and techniques; Radio frequency; Relays; Thin wall structures; Three-term control;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 1996 Conference on
  • Conference_Location
    Braunschweig, Germany
  • Print_ISBN
    0-7803-3376-4
  • Type

    conf

  • DOI
    10.1109/CPEM.1996.547162
  • Filename
    547162