• DocumentCode
    2459559
  • Title

    Microcalorimeter for 3.5 mm 50 /spl Omega/ coaxial transmission line

  • Author

    Janik, D.

  • Author_Institution
    Physikalisch Tech. Bundesanstalt, Braunschweig, Germany
  • fYear
    1996
  • fDate
    17-21 June 1996
  • Firstpage
    456
  • Lastpage
    457
  • Abstract
    At PTB a microcalorimeter has been installed for calibrating RF power mounts with 3.5 mm precision 50 /spl Omega/ coaxial connectors (PC-3.5) to be used as power standards up to 26.5 GHz. Two detachable inserts with 3.5 mm coaxial transmission lines for a twin-type microcalorimeter have been developed. Measurements of bolometer mounts fitted with PC-3.5 adaptors and of a PC-3.5 power sensor are described.
  • Keywords
    calibration; electric connectors; measurement standards; microwave measurement; power measurement; 0 to 26.5 GHz; 3.5 mm; 50 ohm; PTB; RF power mounts; bolometer mounts; calibration; coaxial connectors; coaxial transmission line; detachable inserts; power sensor; power standards; twin-type microcalorimeter; Coaxial components; Conductors; Frequency measurement; Power measurement; Power transmission lines; Radio frequency; Sensor phenomena and characterization; Thermal conductivity; Transmission line measurements; Transmission lines;