• DocumentCode
    2460702
  • Title

    A high accuracy reactive power and energy meter insensitive to frequency variations

  • Author

    Djokic, B. ; Bosnjakovic, P.

  • Author_Institution
    Inst. for Nat. Meas. Stand., Nat. Res. Council of Canada, Ottawa, Ont., Canada
  • fYear
    1996
  • fDate
    17-21 June 1996
  • Firstpage
    482
  • Lastpage
    483
  • Abstract
    The application of a frequency-controlled power-to-pulse-rate conversion in a high accuracy instrument for reactive power and energy measurement at power frequencies is described. Test results show that the meter accuracy is better than 100 ppm regardless of frequency variation of /spl plusmn/5% around nominal frequency.
  • Keywords
    measurement errors; measurement standards; power measurement; power meters; reactive power; energy meter; frequency variations; frequency-controlled power-to-pulse-rate conversion; power frequencies; reactive power meter; Circuit testing; Energy measurement; Frequency conversion; Instruments; Performance evaluation; Phase shifters; Power measurement; Reactive power; Voltage; Watthour meters;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 1996 Conference on
  • Conference_Location
    Braunschweig, Germany
  • Print_ISBN
    0-7803-3376-4
  • Type

    conf

  • DOI
    10.1109/CPEM.1996.547176
  • Filename
    547176