DocumentCode
2461052
Title
Analysis, design and control of a transcutaneous power regulator for artificial heart
Author
Chen, Qianhong ; Wong, Siu Chung ; Tse, Chi K. ; Ruan, Xinbo
Author_Institution
Coll. of Autom. Eng., Nanjing Univ. of Aeronaut.&Astronaut., Nanjing
fYear
2008
fDate
15-19 June 2008
Firstpage
1833
Lastpage
1838
Abstract
Based on commonly used parameters for a generic-transcutaneous transformer model, a remote power supply using resonant topology for artificial heart is analyzed and designed for easy controllability and high efficiency. Primary and secondary windings of the transcutaneous transformer are positioned outside and inside human body respectively for energy transfer. The two large leakage inductances and the mutual inductance of the transformer are varying parameters of the coupling-coefficient which varies with transformer alignment and gap due to external positioning. Varying resonant-frequency resonant-tank circuits are formed using the transformer inductors and external capacitors to obtain a load insensitive frequency for the voltage transfer function at given range of coupling coefficients and loads. Previous researches usually use frequency modulation which may require a wide control frequency range well above the load insensitive frequency. In this paper, fundamental frequency study of the input-to-output voltage transfer function is carried out. Using the proposed control method, the switching frequency can be locked at just above the load insensitive frequency at heavy load for best efficiency. Specifically, above resonant operation in driving the resonant circuits when varying the coupling-coefficient is maintained using a digital-phase-lock-loop (PLL) technique to achieve zero-voltage switching of a full-bridge switches configuration which is also programmed to provide pulse-width-modulation (PWM) in controlling the output voltage. A prototype transcutaneous power regulator is built and found to have good efficiency and regulation in responding to changing alignment or gap of the transcutaneous transformer, load and input voltage dynamically.
Keywords
artificial organs; digital phase locked loops; switching convertors; transfer functions; transformer windings; transformers; zero voltage switching; artificial heart; digital-phase-lock-loop technique; external capacitors; frequency modulation; full-bridge switches configuration; generic transcutaneous transformer model; leakage inductances; mutual inductance; pulse-width-modulation; remote power supply; resonant topology; resonant-frequency resonant-tank circuits; transcutaneous power regulator; voltage transfer function; zero- voltage switching; Artificial heart; Biological system modeling; Coupling circuits; Frequency; Pulse width modulation; RLC circuits; Regulators; Resonance; Transfer functions; Zero voltage switching;
fLanguage
English
Publisher
ieee
Conference_Titel
Power Electronics Specialists Conference, 2008. PESC 2008. IEEE
Conference_Location
Rhodes
ISSN
0275-9306
Print_ISBN
978-1-4244-1667-7
Electronic_ISBN
0275-9306
Type
conf
DOI
10.1109/PESC.2008.4592210
Filename
4592210
Link To Document