DocumentCode :
2462163
Title :
A micropotentiometer-based system for low-voltage calibration of alternating voltage measurement standards
Author :
Budovsky, I.
Author_Institution :
Nat. Meas. Lab., CSIRO, Lindfield, NSW, Australia
fYear :
1996
fDate :
17-21 June 1996
Firstpage :
497
Lastpage :
498
Abstract :
The paper describes a new ac-dc transfer standard developed at the National Measurement Laboratory. It enables calibration of ac-dc transfer and absolute ac voltage standards to be made in the range of voltages from 2 mV to 500 mV, with an uncertainty of 10 to 1000 ppm in the frequency range 10 Hz to 1 MHz.
Keywords :
calibration; measurement standards; potentiometers; voltage measurement; 10 Hz to 1 MHz; 2 to 500 mV; National Measurement Laboratory; absolute ac voltage standards; ac-dc transfer; alternating voltage measurement standards; low-voltage calibration; measurement uncertainty; micropotentiometer; Australia; Calibration; Frequency; Laboratories; Measurement standards; Physics; Resistors; Standards development; Thermal resistance; Voltage measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 1996 Conference on
Conference_Location :
Braunschweig, Germany
Print_ISBN :
0-7803-3376-4
Type :
conf
DOI :
10.1109/CPEM.1996.547184
Filename :
547184
Link To Document :