A micropotentiometer-based system for low-voltage calibration of alternating voltage measurement standards
Author
Budovsky, I.
Author_Institution
Nat. Meas. Lab., CSIRO, Lindfield, NSW, Australia
fYear
1996
fDate
17-21 June 1996
Firstpage
497
Lastpage
498
Abstract
The paper describes a new ac-dc transfer standard developed at the National Measurement Laboratory. It enables calibration of ac-dc transfer and absolute ac voltage standards to be made in the range of voltages from 2 mV to 500 mV, with an uncertainty of 10 to 1000 ppm in the frequency range 10 Hz to 1 MHz.
Keywords
calibration; measurement standards; potentiometers; voltage measurement; 10 Hz to 1 MHz; 2 to 500 mV; National Measurement Laboratory; absolute ac voltage standards; ac-dc transfer; alternating voltage measurement standards; low-voltage calibration; measurement uncertainty; micropotentiometer; Australia; Calibration; Frequency; Laboratories; Measurement standards; Physics; Resistors; Standards development; Thermal resistance; Voltage measurement;