DocumentCode :
2462392
Title :
Evaluation of the calculable high frequency ac/dc standard
Author :
vanMullem, C.J. ; Janssen, W.J.G.D. ; deVreede, J.P.M.
Author_Institution :
NMi van Swinden Lab., Delft, Netherlands
fYear :
1996
fDate :
17-21 June 1996
Firstpage :
499
Lastpage :
500
Abstract :
The calculable HF ac/dc transfer standard has been developed at VSL. An improved model which contains a more detailed description of the thermal element with the end disc and the housing, is presented together with measurements. The major result is a calculated shift of about 4000 ppm in the ac/dc difference at 100 MHz. Based on this model, a 2 V standard is under development.
Keywords :
measurement standards; voltage measurement; 100 MHz; 2 V; HF; VSL; ac/dc standard; end disc; housing; modelling; thermal element; transfer standard; Capacitance; Conducting materials; Conductivity; Frequency measurement; Measurement standards; Parameter estimation; Predictive models; Tellurium; Uncertainty; Wire;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 1996 Conference on
Conference_Location :
Braunschweig, Germany
Print_ISBN :
0-7803-3376-4
Type :
conf
DOI :
10.1109/CPEM.1996.547185
Filename :
547185
Link To Document :
بازگشت