Modeling cardiac pacemaker malfunctions with the Virtual Heart Model
Author
Jiang, Zhihao ; Mangharam, Rahul
Author_Institution
Department of Electrical and System Engineering, University of Pennsylvania
fYear
2011
fDate
Aug. 30 2011-Sept. 3 2011
Firstpage
263
Lastpage
266
Abstract
Implantable cardiac devices such as artificial pacemakers deliver therapies according to the timing information from the heart. Such devices work under the assumptions of perfect sensing, which are: (a) the pacemaker leads remain in place, and (b) the pacing therapy in one chamber (e.g. atrium) is insulated from the other chambers (e.g. ventricles). But there are common cases which violate these assumptions and the mechanisms for imperfect sensing cannot be captured by a simple signal generator. In this paper we use the Penn Virtual Heart Model (VHM) to investigate the spatial and temporal aspects of the electrical conduction system of the heart in a closed-loop with a pacemaker model. We utilize the spatial properties of the heart to model the sensing mechanism, and use clinical cases to show the validity of our sensing model. Such closed-loop evaluation of the pacemaker operation allows for functional testing of pacemaker software, the development of new algorithms for rhythm therapy and also serves as a tool for incoming cardiac electrophysiology fellows.
Keywords
Automata; Electrocardiography; Heart; Lead; Pacemakers; Sensors; Timing; Computer Simulation; Equipment Failure; Equipment Failure Analysis; Heart Conduction System; Humans; Models, Cardiovascular; Pacemaker, Artificial;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
Conference_Location
Boston, MA
ISSN
1557-170X
Print_ISBN
978-1-4244-4121-1
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2011.6090051
Filename
6090051
Link To Document