Title :
Ablation of demineralized dentin using a mid-infrared tunable nanosecond pulsed laser at 6 μm wavelength range for selective excavation of carious dentin
Author :
Ishii, K. ; Saiki, M. ; Yoshikawa, K. ; Yasuo, K. ; Yamamoto, K. ; Awazu, K.
Author_Institution :
Graduate School of Engineering, Osaka University, Osaka 565-0771 Japan
fDate :
Aug. 30 2011-Sept. 3 2011
Abstract :
In dental clinic, some lasers have already realized the optical drilling of dental hard tissue. However, conventional lasers lack the ability to discriminate and excavate carious tissue only, and still depend on the dentist´s ability. The objective of this study is to develop a selective excavation of carious dentin by using the laser ablation with 6 μm wavelength range. Bovine dentin demineralized with lactic acid solution was used as a carious dentin model. A mid-infrared tunable pulsed laser was obtained by difference-frequency generation technique. The wavelength was tuned around the absorption bands called amide 1 and amide 2. In the wavelength range from 5.75 to 6.60 μm, the difference of ablation depth between demineralized and normal dentin was observed. The wavelength at 6.02 μm and the average power density of 15 W/cm2, demineralized dentin was removed selectively with less-invasive effect on normal dentin. The wavelength at 6.42 μm required the increase of average power density, but also showed the possibility of selective ablation. In the near future, development of compact laser device will open the minimal invasive laser treatment to the dental clinic.
Keywords :
Absorption; Dentistry; Educational institutions; Laser ablation; Measurement by laser beam; Radiation effects; Semiconductor lasers; Animals; Cattle; Dental Caries; Dentin; Laser Therapy; Treatment Outcome;
Conference_Titel :
Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
Conference_Location :
Boston, MA
Print_ISBN :
978-1-4244-4121-1
Electronic_ISBN :
1557-170X
DOI :
10.1109/IEMBS.2011.6090083