  • Abstract
    A formerly developed ultrasound liver phantom for testing of surgical navigation systems and liver resection trainings was evaluated experimentally. The phantom was scanned with CT and the dataset was analyzed with existing segmentation techniques. A virtual 3D model was generated on the basis of the segmentation; it was later used for phantom registration in a surgical assistance navigation system. Within an experiment, ten test persons have tried to touch three tumor models hidden in the phantom with the tip of a resection instrument. In 67% of overall 30 touch trials it was a successful touch at the first go. It means that the developed liver phantom is appropriate for testing of surgical navigation systems, as well as for computer assisted liver resection trainings.
  • Keywords
    Computed tomography; Instruments; Liver; Phantoms; Surgery; Tumors; Ultrasonic imaging; Equipment Design; Equipment Failure Analysis; Hepatectomy; Humans; Liver Neoplasms; Phantoms, Imaging; Reproducibility of Results; Sensitivity and Specificity; Surgery, Computer-Assisted; Tomography, X-Ray Computed; Ultrasonography, Interventional;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
  • Conference_Location
    Boston, MA
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-4121-1
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2011.6090205
  • Filename
    6090205