DocumentCode :
2468585
Title :
A radial filtering scheme applied to intracoronary ultrasound images
Author :
Van Horn, Mark H. ; Snyder, Wesley E. ; Herrington, David M.
Author_Institution :
Bowman Gray Sch. of Med., Winston-Salem, NC, USA
fYear :
1993
fDate :
5-8 Sep 1993
Firstpage :
117
Lastpage :
120
Abstract :
Intracoronary ultrasound (ICUS) is a new method to quantify the degree and severity of coronary atherosclerosis. The accurate, precise, and preferably automated detection of lumen-intimal and medial-adventitial boundaries is desirable. However, because of the poor quality of ICUS images, determining accurate boundary estimates is often difficult. Low contrast, artifacts, and low signal-to-noise ratios affect both the quality of the image and the accuracy of boundary defection. Using synthetic image data, the accuracy of two different implementations of a radially oriented edge detection operator were evaluated. The results demonstrate that the application of a radially oriented directional derivative operator in conjunction with a Gaussian blurring function can be used to generate accurate lumen-intimal and medial-adventitial boundary estimates in ICUS images
Keywords :
biomedical ultrasonics; cardiology; edge detection; medical image processing; Gaussian blurring function; artifacts; automated detection; coronary atherosclerosis quantification; image quality; intracoronary ultrasound images; low contrast; lumen-intimal boundary; medial-adventitial boundary; medical diagnostic imaging; radial filtering scheme; radially oriented directional derivative operator; signal-to-noise ratio; Biomedical imaging; Catheters; Convolution; Data mining; Filtering; Image edge detection; Impedance; Signal to noise ratio; Smoothing methods; Ultrasonic imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computers in Cardiology 1993, Proceedings.
Conference_Location :
London
Print_ISBN :
0-8186-5470-8
Type :
conf
DOI :
10.1109/CIC.1993.378490
Filename :
378490
Link To Document :
بازگشت