DocumentCode :
2471741
Title :
Implementation and validation of a power-efficient, high-speed modulation design for wireless oxygen saturation measurement systems
Author :
Shaltis, Phillip ; Sokwoo Rhee ; Asada, Haruhiko H.
Author_Institution :
Dept. of Mech. Eng., MIT, Cambridge, MA, USA
fYear :
2002
fDate :
2002
Firstpage :
193
Lastpage :
194
Abstract :
This paper describes the implementation and initial validation results of a novel power-efficient, high-speed modulation design for wearable oxygen saturation sensor systems. The research presented is in conjunction with the development of continuous photo plethysmographic health monitoring device known as the Ring Sensor. It is demonstrated that a high LED modulation rate coupled with a low duty cycle significantly reduce LED power consumption. Additionally, initial benchmarking results using a Nellcor N-395 pulse oximeter indicate good agreement with the new design for both heart rate and oxygen saturation measurements
Keywords :
biomedical electronics; biomedical telemetry; fast Fourier transforms; light emitting diodes; medical signal processing; modulation; oximetry; patient monitoring; signal sampling; FFT; LED power consumption; Nellcor pulse oximeter; Ring Sensor; continuous photoplethysmographic monitoring; high-speed modulation design; power spectrum; power-efficient design; signal stability; wearable sensor systems; wireless oxygen saturation measurement systems; Delay; Energy consumption; Fingers; Heart rate; Light emitting diodes; Monitoring; Oxygen; Photodetectors; Pulse measurements; Sampling methods;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Bioengineering Conference, 2002. Proceedings of the IEEE 28th Annual Northeast
Conference_Location :
Philadelphia, PA
Print_ISBN :
0-7803-7419-3
Type :
conf
DOI :
10.1109/NEBC.2002.999531
Filename :
999531
Link To Document :
بازگشت