Abstract :
This paper compares the acoustic-beam properties of high-frequency linear and annular arrays operating at 40 MHz. The annular array was modeled with a spatial-impulse response method and the linear array was simulated with Field II. The parameters of the linear array were 64-mum pitch, 14-mum kerf, 3-mm elevation aperture, 12-mm elevation focus and up to 64 elements. The parameters of our 40-MHz annular array were 5 elements, 12-mm geometric focus, and 6-mm total aperture. Simulations were performed for the cases of fixed, synthetic- transmit/receive, and dynamic-receive focusing. After synthetic- focusing, the -6-dB axial depth of field (DOF) for the annular array was ap 5 mm, while the DOF of the linear array was 4 mm. Over the extended DOF of 9 to 15 mm, the azimuthal beamwidths of the annular array ranged from 65 to 89 mum, while the beamwidths of the linear array ranged from 110 to 180 mum. The practical implication of the above analysis is that an annular array will lead to better overall image quality than current state-of-the-art in HFU linear-array technology. In vivo mouse embryo data were acquired with the 40-MHz annular array and beamformed in order to demonstrate the improved image quality obtained with synthetic focusing versus fixed and dynamic-receive focusing.
Keywords :
biomedical transducers; biomedical ultrasonics; ultrasonic focusing; ultrasonic transducer arrays; HFU technology; acoustic beam properties; dynamic-receive focusing; fixed focusing; frequency 40 MHz; high-frequency annular arrays; high-frequency linear arrays; image quality; in vivo mouse embryo data; spatial-impulse response method; synthetic focusing; Acoustic arrays; Acoustic beams; Apertures; Array signal processing; Connectors; Focusing; Image quality; Microwave antenna arrays; Ultrasonic imaging; Ultrasonic transducers;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 2007. IEEE
Conference_Location :
New York, NY
ISSN :
1051-0117
Print_ISBN :
978-1-4244-1383-6
Electronic_ISBN :
1051-0117
Type :
conf
DOI :
10.1109/ULTSYM.2007.425
Filename :
4409998
Link To Document :
بازگشت