DocumentCode :
2481326
Title :
Multi-channel cross-correlation for increasing sensitivity in voltage noise measurements
Author :
Scandurra, G. ; Ciofi, C.
Author_Institution :
Dip. Fis. della Materia e Ing. Elettron., Univ. of Messina, Messina, Italy
fYear :
2012
fDate :
13-16 May 2012
Firstpage :
1524
Lastpage :
1528
Abstract :
In this paper we demonstrate that if N (N>;2) voltage amplifiers are connected to the device under test when performing voltage noise measurements in a situation in which the equivalent voltage noise of each single amplifier is larger than that to be measured, the time required for obtaining a given accuracy is reduced by a factor N(N-1)/2 with respect to the conventional set-up employing just two amplifiers. Measurements on two different set-ups employing up to 4 channels and up to 8 channels are presented that confirm that the expected reduction factors in measurement time by a factor of 6 and 28, respectively, are indeed obtained in actual experiments.
Keywords :
amplifiers; electric noise measurement; voltage measurement; device under test; equivalent voltage noise measurement; multichannel cross-correlation; reduction factor; time measurement; voltage amplifier; Bandwidth; Correlation; Noise; Noise measurement; Preamplifiers; Time measurement; Voltage measurement; Noise measurements; cross-correlation; spectral analysis;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Instrumentation and Measurement Technology Conference (I2MTC), 2012 IEEE International
Conference_Location :
Graz
ISSN :
1091-5281
Print_ISBN :
978-1-4577-1773-4
Type :
conf
DOI :
10.1109/I2MTC.2012.6229441
Filename :
6229441
Link To Document :
بازگشت