• DocumentCode
    2481679
  • Title

    A MICS telemetry implant powered by a 27MHz ISM inductive link

  • Author

    Laskovski, A.N. ; Yuce, M.R.

  • Author_Institution
    Sch. of Electr. Eng. & Comput. Sci., Univ. of Newcastle, Callaghan, NSW, Australia
  • fYear
    2011
  • fDate
    Aug. 30 2011-Sept. 3 2011
  • Firstpage
    2909
  • Lastpage
    2912
  • Abstract
    An implantable telemetry device is presented, inductively powered by a 5.8mW 27MHz ISM-band supply and received with a stacked-spiral implantable power coil. A 15th harmonic 405MHz MICS-band data carrier is generated from the power signal, data-modulated, filtered and sent to a receiver via a wireless data link. The implantable module is tested in a biological environment, successfully proving the concept.
  • Keywords
    biomedical electronics; biomedical telemetry; coils; modules; patient monitoring; prosthetics; ISM inductive link; MICS telemetry implant; frequency 27 MHz; frequency 405 MHz; implantable module; power 5.8 mW; stacked-spiral implantable power coil; Coils; Harmonic analysis; Implants; Microwave integrated circuits; Spirals; Telemetry; Wireless communication; Radio Waves; Telemetry;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
  • Conference_Location
    Boston, MA
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-4121-1
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2011.6090801
  • Filename
    6090801