• DocumentCode
    2482123
  • Title

    Fascicle-selective multi-contact cuff electrode

  • Author

    Rossel, O. ; Soulier, Fabien ; Coulombe, Jonathan ; Bernard, Sebastien ; Cathebras, Guy

  • Author_Institution
    Lab. d´Inf. de Robot. et de Microelectron. de Montpellier, Montpellier, France
  • fYear
    2011
  • fDate
    Aug. 30 2011-Sept. 3 2011
  • Firstpage
    2989
  • Lastpage
    2992
  • Abstract
    Neural recording is one of the main issues to be addressed in order to allow closed-loop functional electrical stimulation systems. Because each fascicle in nerves carry specific information, new sensors providing high spatial selectivity are required for chronic implantable devices. This work aims at evaluating the feasibility of a new device using a highly spatial-selective multi-contact cuff electrode. The proposed electrode configuration is evaluated based on simulations using a model of a nerve comprising multiple fascicles. Study of the electrode selectivity is done and compared with a state-of-the-art electrode designed for the same purpose and shows that activity of two fascicles separated by as little as 1 mm can be distinguished. Implementation challenges and perspectives for such electrodes are also discussed.
  • Keywords
    biomedical electrodes; neuromuscular stimulation; neurophysiology; chronic implantable devices; electrode selectivity; fascicle-selective multicontact cuff electrode; functional electrical stimulation; neural recording; state-of-the-art electrode; Analytical models; Compounds; Electrodes; Extracellular; Nerve fibers; Shape; Silicon; Electrodes; Feasibility Studies;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
  • Conference_Location
    Boston, MA
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-4121-1
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2011.6090820
  • Filename