Abstract :
A step down procedure is described for establishing voltage standards in the 1 mV to 100 mV range at frequencies between 10 Hz and 1 MHz. The step down employs low voltage thermal voltage converters and micropotentiometers. Techniques are given for measuring input impedance and calculating loading errors.
Keywords :
measurement errors; measurement standards; potentiometers; voltage measurement; 1 to 100 mV; 10 Hz to 1 MHz; loading errors; low voltage standards; low voltage thermal voltage converters; measuring input impedance; micropotentiometers; voltage standards; Capacitance; Digital-to-frequency converters; Electric resistance; Electrical resistance measurement; Frequency conversion; Impedance measurement; Low voltage; NIST; Signal generators; Uncertainty;