• DocumentCode
    2486817
  • Title

    Time-domain measurement method to guard against preamplifier saturation

  • Author

    Jackson, Matthew J.

  • Author_Institution
    EMC Lab., DENSO Int. America, Inc., Southfield, MI, USA
  • fYear
    2010
  • fDate
    25-30 July 2010
  • Firstpage
    479
  • Lastpage
    483
  • Abstract
    This paper investigates the cause of preamplifier saturation, discusses the 1 dB compression point and considers the ramifications of collecting data with a saturated amplifier. Two common saturation analysis methods, the attenuator check and band-pass filter, are considered as ways to guard against preamplifier saturation. After a brief discussion on the amplifier´s input signal and its composition, a time-domain measurement method, consisting of seven steps, is introduced as another way to prevent preamplifier saturation.
  • Keywords
    band-pass filters; preamplifiers; time measurement; time-domain analysis; attenuator check; bandpass filter; preamplifier saturation prevention; saturated amplifier; saturation analysis methods; time-domain measurement method; Antenna measurements; Band pass filters; Gain; Oscilloscopes; Preamplifiers; Time domain analysis; Voltage measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electromagnetic Compatibility (EMC), 2010 IEEE International Symposium on
  • Conference_Location
    Fort Lauderdale, FL
  • ISSN
    2158-110X
  • Print_ISBN
    978-1-4244-6305-3
  • Type

    conf

  • DOI
    10.1109/ISEMC.2010.5711322
  • Filename
    5711322