DocumentCode
2486961
Title
Internal evaluation of registration results for radiographic images
Author
Selby, Boris Peter ; Sakas, Georgios ; Walter, Stefan ; Groch, Wolf-Dieter ; Stilla, Uwe
Author_Institution
Medcom GmbH, Darmstadt
fYear
2008
fDate
8-11 Dec. 2008
Firstpage
1
Lastpage
4
Abstract
This wort focuses on internal gray level based evaluation of image registration results. The motivation is to provide all approach for self-diagnosis in the scope of a patient alignment system based on rigid registration of real and reconstructed X-ray images. As an automatic system should provide expressive indicators for the correctness of the outcome, we propose a method to estimate the probability for the resulting transformations to lie within a predefined window of acceptable values. Based purely on image gray values, the approach is independent from previous knowledge about the images. By registration of corresponding fragments of both images we generate redundancy and define the probability density of the resulting transformations. The proposed method is tested comparing digital reconstructed radiographs (DRRs) to X-ray images. By introducing geometric and radiometric deviations we show that a reliable self-diagnosis is possible.
Keywords
geometry; image reconstruction; image registration; medical image processing; radiography; X-ray image reconstruction; digital reconstructed radiographs; geometric-radiometric deviations; image gray values; patient alignment system; probability density; registration internal evaluation; Biomedical imaging; Image generation; Image reconstruction; Image registration; Neural networks; Probability; Radiography; Redundancy; Testing; Tiles;
fLanguage
English
Publisher
ieee
Conference_Titel
Pattern Recognition, 2008. ICPR 2008. 19th International Conference on
Conference_Location
Tampa, FL
ISSN
1051-4651
Print_ISBN
978-1-4244-2174-9
Electronic_ISBN
1051-4651
Type
conf
DOI
10.1109/ICPR.2008.4761692
Filename
4761692
Link To Document