DocumentCode
2488104
Title
Measurement of respiration rate in preterm infants by laser Doppler vibrometry
Author
Scalise, Lorenzo ; Ercoli, Ilaria ; Marchionni, Paolo ; Tomasini, Enrico Primo
Author_Institution
Dipt. di Meccanica, Univ. Politec. delle Marche, Ancona, Italy
fYear
2011
fDate
30-31 May 2011
Firstpage
657
Lastpage
661
Abstract
The continuous monitoring of patients recovered in intensive care units is a mandatory tasks for clinicians in order to check conditions and effects of therapies. Among the many signals and quantities observable, the ones related to the heart and particularly the lung´s activity are of primary interest. Today the majority of the measurement methods used for the assessment of the respiration rate are based on invasive methods (spriometry) which are often included into mechanical ventilator to which the patients, in need of assisted breathing, are attached. In this paper an optical, non-contact measurement method for monitoring of the respiration rate in pre-term infants is reported. It is based on the assessment of abdominal wall movements detected by a laser Doppler vibrometer and can be operated at a distance from the patient and outside the incubator. A total of 9 preterm patients have been monitored; respiration rates measured by a reference instrument (BabyLog 8000 Draeger plus) and the proposed non-contact method have been measured for a period of time of 60s. Data have been compared using the Bland-Altman analysis and results demonstrate an optimal correlation (R=0.97) among data and an uncertainty of the proposed method of ± 0.13s (K=2).
Keywords
laser applications in medicine; paediatrics; patient monitoring; pneumodynamics; vibration measurement; BabyLog 8000 Draeger plus; Bland-Altman analysis; abdominal wall movement; continuous patient monitoring; intensive care unit; laser Doppler vibrometry; mechanical ventilator; preterm infant; respiration rate measurement; Instruments; Measurement by laser beam; Monitoring; Optical fibers; Optical sensors; Pediatrics; Skin; laser doppler vibrometry; non contact; respiration rate;
fLanguage
English
Publisher
ieee
Conference_Titel
Medical Measurements and Applications Proceedings (MeMeA), 2011 IEEE International Workshop on
Conference_Location
Bari
Print_ISBN
978-1-4244-9336-4
Type
conf
DOI
10.1109/MeMeA.2011.5966740
Filename
5966740