Title :
A continuous, wearable, and wireless heart monitor using head ballistocardiogram (BCG) and head electrocardiogram (ECG)
Author :
He, David Da ; Winokur, Eric S. ; Sodini, Charles G.
Author_Institution :
Dept. of Electr. Eng. & Comput. Sci., Massachusetts Inst. of Technol., Cambridge, MA, USA
fYear :
2011
fDate :
Aug. 30 2011-Sept. 3 2011
Firstpage :
4729
Lastpage :
4732
Abstract :
Continuous and wearable heart monitoring is essential for early detection and diagnosis of cardiovascular diseases. We demonstrate a continuous, wearable, and wireless heart monitor that is worn at the ear. The device has the form factor of a hearing aid and is wirelessly connected to a PC for data recording and analysis. With the ear as an anchoring point, the heart monitor measures the ballistocardiographic (BCG) motion of the head using a MEMS tri-axial accelerometer, which is an electrode-less method to measure heart rate. Additionally, electrocardiogram (ECG) is measured locally near the ear using a single-lead configuration. The peak timing delay between the head ECG and the head BCG, or RJ interval, can be extracted in the presence of noise using cross-correlation. The RJ interval is shown to correlate to the heart´s pre-ejection period during both Valsalva and whole-body tilt maneuvers.
Keywords :
accelerometers; bioMEMS; biomedical equipment; data analysis; data recording; diseases; ear; electrocardiography; patient monitoring; MEMS triaxial accelerometer; PC; Valsalva tilt maneuvers; ballistocardiographic motion; cardiovascular diseases; data analysis; data recording; ear; electrode-less method; head BCG; head ECG; head ballistocardiogram; head electrocardiogram; hearing aid; heart rate; noise; single-lead configuration; timing delay; wearable heart monitor; whole-body tilt maneuvers; wireless heart monitor; Biomedical monitoring; Ear; Electrocardiography; Heart; Magnetic heads; Monitoring; Noise; Electrocardiography; Equipment Design; Head; Heart; Hemodynamics; Humans; Monitoring, Physiologic; Radio Waves;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
Conference_Location :
Boston, MA
ISSN :
1557-170X
Print_ISBN :
978-1-4244-4121-1
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/IEMBS.2011.6091171