Title :
Dual inductive link coil design for a neural recording system
Author :
Rush, Alexander ; Troyk, Philip R.
Author_Institution :
Illinois Inst. of Technol., Chicago, IL, USA
fDate :
Aug. 30 2011-Sept. 3 2011
Abstract :
This paper reports an approach to the physical design of the coils used in a dual inductive link to provide two-way wireless communication and power for a neural recording system. The design approach makes use of an analytic model of the link performance in terms of the physical parameters of the link, which allows physical parameters to be iterated on a computer rather than on the bench to find the optimal design within the physical restrictions imposed. In particular, this approach was used to choose the optimal implant data coil sizing to maximize the difference between the contributions of the constructive and destructive paths of the reverse telemetry signal.
Keywords :
biomedical electronics; biomedical telemetry; coils; inductive power transmission; neurophysiology; power supplies to apparatus; prosthetics; wireless sensor networks; analytic model; coil physical design; dual inductive link coil design; link performance; link physical parameters; neural recording system; optimal implant data coil; reverse telemetry signal; two way wireless communication; wireless power; Coils; Frequency measurement; Implants; Inductance; Mathematical model; Telemetry; Wires; Algorithms; Computer Communication Networks; Computer Systems; Electric Capacitance; Electric Power Supplies; Electronics; Equipment Design; Humans; Man-Machine Systems; Models, Statistical; Neurons; Signal Processing, Computer-Assisted; Telemetry; User-Computer Interface; Wireless Technology;
Conference_Titel :
Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
Conference_Location :
Boston, MA
Print_ISBN :
978-1-4244-4121-1
Electronic_ISBN :
1557-170X
DOI :
10.1109/IEMBS.2011.6091579