  • Abstract
    Magnetic Resonance Imaging (MRI) can provide three dimensional (3D) imaging with excellent resolution and sensitivity making it ideal for guiding and monitoring interventions. The development of MRI-compatible interventional devices is complicated by factors including: the high magnetic field strength, the requirement that such devices should not degrade image quality, and the confined physical space of the scanner bore. Numerous MRI guided actuated devices have been developed or are currently being developed utilizing piezoelectric actuators as their primary means of mechanical energy generation to enable better interventional procedure performance. While piezoelectric actuators are highly desirable for MRI guided actuation for their precision, high holding force, and non-magnetic operation they are often found to cause image degradation on a large enough to scale to render live imaging unusable. This paper describes a newly developed piezoelectric actuator driver and control system designed to drive a variety of both harmonic and non-harmonic motors that has been demonstrated to be capable of operating both harmonic and non-harmonic piezoelectric actuators with less than 5% SNR loss under closed loop control. The proposed system device allows for a single controller to control any supported actuator and feedback sensor without any physical hardware changes.
  • Keywords
    biomedical MRI; medical image processing; motion control; piezoelectric actuators; 3D imaging; MRI compatible interventional devices; Magnetic Resonance Imaging; field strength; image quality; mechanical energy; motion control; nonharmonic motors; piezoelectric actuators; sensitivity; Actuators; Harmonic analysis; Magnetic resonance imaging; Robots; Signal to noise ratio; Computers; Electricity; Equipment Design; Humans; Imaging, Three-Dimensional; Magnetic Resonance Imaging; Magnetics; Motion; Phantoms, Imaging; Robotics; Sensitivity and Specificity; Signal-To-Noise Ratio; Software; Stress, Mechanical; Torque;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
  • Conference_Location
    Boston, MA
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-4121-1
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2011.6091649
  • Filename
    6091649