DocumentCode
2500910
Title
Strength evaluation of a variable diameter acetabular trial implant under realistic loading conditions
Author
Levasseur, A. ; Petit, Y. ; Dansereau, M. ; Fernandes, J.C.
Author_Institution
Imaging & Orthopaedic Res. Lab., Montreal, QC, Canada
fYear
2011
fDate
Aug. 30 2011-Sept. 3 2011
Firstpage
7433
Lastpage
7436
Abstract
A variable diameter acetabular trial implant (VDATI) was designed to reduce the costs related to cleaning, sterilisation and storage of surgical instruments used for total hip arthroplasty. The purpose of this study was to evaluate the mechanical strength of a functional prototype of the VDATI. Experimental testing was performed to identify if the VDATI can resist loading conditions similar to the ones occurring during the surgical procedure and to validate a finite element model (FEM) of the VDATI. The results highlighted the potential of the current concept of the VDATI and demonstrated the relevance to continue its development.
Keywords
biomechanics; cleaning; finite element analysis; mechanical strength; prosthetics; sterilisation (microbiological); surgery; cleaning; finite element model; loading conditions; mechanical strength; realistic loading conditions; sterilisation; strength evaluation; surgical instruments; total hip arthroplasty; variable diameter acetabular trial implant; Finite element methods; Hip; Loading; Prototypes; Stress; Surgery; Testing; Acetabulum; Arthroplasty, Replacement, Hip; Humans; Materials Testing; Stress, Mechanical; Weight-Bearing;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
Conference_Location
Boston, MA
ISSN
1557-170X
Print_ISBN
978-1-4244-4121-1
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2011.6091731
Filename
6091731
Link To Document