DocumentCode :
250116
Title :
Smart Tissue Anastomosis Robot (STAR): Accuracy evaluation for supervisory suturing using near-infrared fluorescent markers
Author :
Leonard, S. ; Shademan, Azad ; Yonjae Kim ; Krieger, Alexandra ; Kim, Peter C. W.
Author_Institution :
Children´s Nat. Med. Center, Sheikh Zayed Inst. for Pediatric Surg. Innovation, Washington, DC, USA
fYear :
2014
fDate :
May 31 2014-June 7 2014
Firstpage :
1889
Lastpage :
1894
Abstract :
This paper specifies and evaluates the accuracy of the Smart Tissue Anastomosis Robot (STAR). The STAR is a proof of concept vision-guided robotic system equipped with an actuated laparoscopic suturing tool and a multispectral vision system. The STAR supports image-based suturing commands and is capable of detecting near-infrared fluorescent (NIRF) markers that provide reliable visual segmentation and tracking. The paper reports the best case scenario accuracy specifications of the STAR as derived from its configuration and calibration parameters. We also evaluate experimentally the effects of overlaying NIRF markers on the accuracy of the STAR when these markers are used as the source of image-based commands and we compare these results to the accuracy of the STAR with image-based commands generated from plain color images. Our results demonstrate that the STAR is able to place sutures on a planar phantom with an average accuracy of 0.5 mm with a standard deviation of 0.2 mm and that NIRF markers have no statistically significant adverse effect on the accuracy.
Keywords :
image colour analysis; image segmentation; medical robotics; object detection; object tracking; robot vision; surgery; NIRF marker detection; STAR; calibration parameter; color images; concept vision-guided robotic system; configuration parameter; image-based commands; image-based suturing command; laparoscopic suturing tool; multispectral vision system; near-infrared fluorescent marker; near-infrared fluorescent markers; smart tissue anastomosis robot; supervisory suturing; visual segmentation; visual tracking; Accuracy; Calibration; Cameras; Robot kinematics; Robot vision systems; Surgery;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Robotics and Automation (ICRA), 2014 IEEE International Conference on
Conference_Location :
Hong Kong
Type :
conf
DOI :
10.1109/ICRA.2014.6907108
Filename :
6907108
Link To Document :
بازگشت