Title
Compensating for torsion windup in steerable needles
Author
Reed, Kyle B.
Author_Institution
Dept. of Mech. Eng., Johns Hopkins Univ., Baltimore, MD
fYear
2008
fDate
19-22 Oct. 2008
Firstpage
936
Lastpage
941
Abstract
Long, flexible, bevel-tip needles curve during insertion into tissue, and rotating the base reorients the tip to steer subsequent insertions. Friction between the tissue and the needle shaft, however, can cause a severe discrepancy between the needle base and tip angles. In this paper, I demonstrate an algorithm to properly align the entire length of the needle using torque measured at the base. My algorithm uses several intermediate base rotations to align the orientation of points along the shaft with the desired angle, with minimal remaining torque exerted by the base. I performed an experimental validation with four angle sensors attached to the needle throughout the tissue. My compensation algorithm decreased the lag throughout the needle by up to 88%.
Keywords
angular measurement; biomedical engineering; medical control systems; needles; position control; torque measurement; torsion; angle sensors; flexible bevel tip needles; needle alignment; steerable needles; tissue-needle shaft friction; torque measurement; torsion windup compensation; Brachytherapy; Current measurement; Frequency; Friction; Needles; Robot sensing systems; Shafts; Torque measurement; USA Councils; Windup;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Robotics and Biomechatronics, 2008. BioRob 2008. 2nd IEEE RAS & EMBS International Conference on
Conference_Location
Scottsdale, AZ
Print_ISBN
978-1-4244-2882-3
Electronic_ISBN
978-1-4244-2883-0
Type
conf
DOI
10.1109/BIOROB.2008.4762825
Filename
4762825
Link To Document