DocumentCode :
2509015
Title :
Finite Element Analysis of LTI Carbon Hip Joint Head
Author :
Chen Lan ; Du Xiuli ; Zheng Yutao ; Bao Yiwang ; Chen Mingzhao ; Hu Ming ; Yu Dan
Author_Institution :
Sch. of Med. Instrum. & Food Eng., Univ. of Shanghai for Sci. & Technol., Shanghai, China
fYear :
2009
fDate :
11-13 June 2009
Firstpage :
1
Lastpage :
4
Abstract :
This paper mainly investigates the carbon material femoral head which is used in the semi-hip replacement. Firstly the 3-D finite element model of the acetabula contacted with the prosthesis of different materials were established and calculate the acetabula´s cotacted stress. Thought comparing the contacting stress of the acetabular, the effect of different material with quite different property is clearly shown. It was found that : stress concentration is much obvious in the contact area of acetabulum. Elastic modulus of the femoral head material has great impact on the acetabula´s stress distribution. The greater of elastic modulus, the bigger of the greatest stress and the more obvious of stress concentration. Therefore, semi-hip replacement with the low elastic modulus carbon femoral head can significantly reduce the acetabular wear and alleviate pain.
Keywords :
biomechanics; biomedical materials; bone; elastic moduli; finite element analysis; orthopaedics; prosthetics; LTI carbon hip joint head; acetabulas cotacted stress; carbon material femoral head; elastic modulus; finite element analysis; prosthesis; semihip replacement; Biological materials; Bones; Carbon; Finite element methods; Head; Hip; Joints; Organic materials; Prosthetics; Stress;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Bioinformatics and Biomedical Engineering , 2009. ICBBE 2009. 3rd International Conference on
Conference_Location :
Beijing
Print_ISBN :
978-1-4244-2901-1
Electronic_ISBN :
978-1-4244-2902-8
Type :
conf
DOI :
10.1109/ICBBE.2009.5162859