DocumentCode
2509163
Title
Four Cylinder Halbach Array for Magnetic Navigation System
Author
Tang Xiao-ying ; Zhao Wei ; Xia Ping-Chou
Author_Institution
Sch. of Inf. Sci. & Technol., Beijing Inst. of Technol., Beijing, China
fYear
2009
fDate
11-13 June 2009
Firstpage
1
Lastpage
4
Abstract
The magnetic navigation system has been widely used for the minimally invasive interventions. The four cylinder Halbach array was introduced into the magnetic navigation system to decrease the high cost. A small model was manufactured and measured. The experiment results indicated that the orientation of magnetic field could be precisely controlled by rotating the magnetization direction of the four cylinder magnets.
Keywords
biomedical engineering; navigation; permanent magnets; surgery; Halbach array; magnetic navigation system; minimally invasive interventions; Biomedical imaging; Catheters; Costs; Engine cylinders; Magnetic fields; Minimally invasive surgery; Navigation; Superconducting magnets; Virtual manufacturing; Wire;