DocumentCode
2510677
Title
Digital Time Alignment of High Resolution PET Inveon Block Detectors
Author
Lenox, M.W. ; Atkins, B.E. ; Pressley, D.R. ; McFarland, A. ; Newport, D.F. ; Siegel, S.B.
Author_Institution
Siemens Molecular Imaging, Knoxville, TN
Volume
4
fYear
2006
fDate
Oct. 29 2006-Nov. 1 2006
Firstpage
2450
Lastpage
2453
Abstract
Time alignment is critical to the operation of high performance PET systems. Optimizing the countrate performance of a given system requires proper timing to minimize the effects of random events as early in the datastream as possible to reduce electronic deadtime. Siemens Inveon PET systems high countrate performance is made possible by their exceedingly tight timing tolerance of 1.22 ns FWHM, allowing the use of an optimal 3.5 ns coincidence window. This paper describes the hardware capabilities used to align the system as well as the processing techniques used to perform the alignment.
Keywords
positron emission tomography; 3.5 ns; PET block detectors; Siemens Inveon; digital time alignment; positron emission tomography; Clocks; Detectors; Hardware; Imaging phantoms; Nuclear and plasma sciences; Photomultipliers; Positron emission tomography; Routing; Time measurement; Timing;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2006. IEEE
Conference_Location
San Diego, CA
ISSN
1095-7863
Print_ISBN
1-4244-0560-2
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2006.354407
Filename
4179521