Title :
Multi-Channel Waveform Sampling ASIC for Animal PET System
Author :
Shimazoe, K. ; Yeom, Jy. ; Takahashi, H. ; Kojo, T. ; Minamikawa, Y. ; Fujita, K. ; Murayama, H.
Author_Institution :
Dept. of Bioeng., Tokyo Univ.
fDate :
Oct. 29 2006-Nov. 1 2006
Abstract :
We have designed and fabricated 2.4 mm by 2.4 mm 2 ch 100 MHz/6 bits waveform sampling front-end (WSFE) ASIC for PET (positron emission tomography) with multiplexer readouts. This chip was designed for GSO-APD gamma-ray detectors and has the function of "waveform sampling" at the speed of 100 MHz. One channel of this chip consists of charge-sensitive preamplifier, VGA (variable gain amplifier), folding-ADC and digital readout circuit and converts the waveform of detector signal to digital values at the speed of ~100 MHz. This digitization in early stage enables flexible digital processing to get timing and energy information of gamma-rays which is indispensable for PET system. Digital circuit has three ways to read signals out, parallel readout, 32 words FIFO memory and 2 to 1 multiplexer. The 2 to 1 multiplexer reduces the number of ASIC pins half by using fast clock with double frequency and enables to include more channels in a chip. We designed new 18 channel WSFE ASIC which includes 2 to 1 multiplexer, 3 to 1 multiplexer and 6 to 1 multiplexer. These multiplexers will greatly decrease the number of pins and realize high integration. This chip will be used for APD-based DOI PET system which requires multi-channel ASIC.
Keywords :
application specific integrated circuits; biomedical electronics; gamma-ray detection; nuclear electronics; positron emission tomography; solid scintillation detectors; 100 MHz; 2.4 mm; FIFO memory; GSO-APD gamma-ray detectors; VGA; animal PET system; application specific integrated circuits; charge-sensitive preamplifier; digital readout circuit; multichannel waveform sampling; multiplexer readouts; positron emission tomography; variable gain amplifier; waveform sampling front-end ASIC; Animals; Application specific integrated circuits; Gain; Gamma ray detectors; Multiplexing; Pins; Positron emission tomography; Preamplifiers; Sampling methods; Signal sampling;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2006. IEEE
Conference_Location :
San Diego, CA
Print_ISBN :
1-4244-0560-2
Electronic_ISBN :
1095-7863
DOI :
10.1109/NSSMIC.2006.354412