DocumentCode :
2510775
Title :
A GATE Monte Carlo Simulation of the Performance of a High-Sensitivity and High-Resolution LSO Based Small Animal PET Camera
Author :
Baghaei, Hossain ; Zhang, Yuxuan ; Li, Hongdi ; Wang, Yu ; Kim, Soonseok ; Ramirez, Rocio ; Liu, Jiguo ; Liu, Shitao ; Wong, Wai-Hoi
Author_Institution :
M. D. Anderson Cancer Center, Texas Univ., Houston, TX
Volume :
4
fYear :
2006
fDate :
Oct. 29 2006-Nov. 1 2006
Firstpage :
2476
Lastpage :
2479
Abstract :
A Monte Carlo simulation of the performance of a high-sensitivity and high-resolution small animal PET scanner with large axial field-of-view (AFOV) is presented. The simulated camera is based on the photomultiplier-quadrant-sharing technique and composed of 180 blocks of 14 times 14 LSO crystals (each crystal is 1.16 times 1.27 times 10 mm3). The designed camera consists of 84 detector rings with an AFOV of 11.6 cm and ring diameter of 16.6 cm. For simulation, we used GATE (Geant4 Application for Tomographic Emission) simulation package. GATE is a GEANT4 based software toolkit for realistic simulation of PET and SPECT systems. In this study, we first validated GATE by comparing its prediction for spatial resolution, absolute sensitivity, NEC, and images of the Data Spectrum Micro Deluxe phantom with the measured data using an existing BGO small animal PET scanner which is also based on the photomultiplier-quadrant-sharing technique and has similar AFOV and ring diameter. Spatial resolution simulation was done for point sources located at the center of the camera and also off the center. Simulation data suggest that new LSO scanner could have a reconstructed radial (tangential) spatial resolution of 1.14 (1.14) mm, 1.31 (1.32) mm, 1.54 (1.52) mm, 2.01 (1.8) mm at the center, and 1 cm, 2 cm and 3 cm off the center, respectively. It predicts that 1.2 mm hot rod inserts in the Micro Deluxe phantom will be distinguishable. Simulation predicts an absolute sensitivity of about 6.
Keywords :
Monte Carlo methods; positron emission tomography; 16.6 cm; Data Spectrum Micro Deluxe phantom; GATE simulation package; Geant4 Application for Tomographic Emission; Monte Carlo simulation; photomultiplier-quadrant-sharing technique; positron emission tomography; small animal PET camera; Animals; Application software; Cameras; Crystals; Detectors; Imaging phantoms; Packaging; Positron emission tomography; Software tools; Spatial resolution;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2006. IEEE
Conference_Location :
San Diego, CA
ISSN :
1095-7863
Print_ISBN :
1-4244-0560-2
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2006.354413
Filename :
4179527
Link To Document :
بازگشت