• DocumentCode
    2513605
  • Title

    Optical Determination of the Cross Talk of CT Detection Systems

  • Author

    Buchinsky, Oded ; Gregorian, Lev ; Uman, Igor ; Wainer, Naor

  • Author_Institution
    Philips Med. Syst. Technol. Ltd., Haifa
  • Volume
    5
  • fYear
    2006
  • fDate
    Oct. 29 2006-Nov. 1 2006
  • Firstpage
    3115
  • Lastpage
    3118
  • Abstract
    One of the most crucial properties of a CT scanner is its spatial resolution. The cross talk is one of the main parameters to influence this property. In this study we use an optical system to characterize CT detectors and photodiode arrays which are employed in such detectors. The PDA response is measured at each pixel as a function of X, Z, and the location within the array, for photodiode arrays made by several different manufacturers. In addition, the contributions due to the optical cross talk through the optical glue and due to the light penetrating the separators between the scintillator pixels are studied. The overall cross-talk results obtained are then compared to the actual cross talk measured on the CT scanner, yielding valuable information at an early stage of the scanner development.
  • Keywords
    X-ray detection; computerised tomography; optical crosstalk; photodiodes; scintillation counters; CT detection systems; CT detectors; CT scanner spatial resolution; computerized tomography; optical cross talk determination; optical glue; photodiode arrays; scanner development; scintillator; Computed tomography; Detectors; Laser beams; Manufacturing; Optical arrays; Optical crosstalk; Particle separators; Photodiodes; Sensor arrays; Spatial resolution;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2006. IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1095-7863
  • Print_ISBN
    1-4244-0560-2
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2006.356534
  • Filename
    4179691