Title :
Calibration measurement setup for band-selective personal exposure meters
Author :
Lauer, Oliver ; Benedickter, Hansruedi ; Neubauer, Georg ; Roosli, Martin ; Frohlich, Jurg
Author_Institution :
Electromagn. Fields & Microwave Electron., Swiss Fed. Inst. of Technol., Zurich, Switzerland
Abstract :
In this paper a calibration measurement setup for band-selective personal exposure meters is presented. The equipment and the methods of the measurement procedure are described together with the uncertainty assessment of the measurement equipment, the calibration and the measurement procedure. The average measurement error for the reference measurement is ± 0.32 dB, while the expanded uncertainty with a confidence interval of 95% is calculated to 2.5 dB. Furthermore a straight forward calibration method is described and tested using the EME SPY 120 device from Antenessa. Up to seven different measurements must be performed for a full characterization of personal exposure meters. The presented measurement setup leads to a higher accuracy of the calibration and the achieved results will allow for an improved exposure assessment in experimental and epidemiological studies.
Keywords :
anechoic chambers (electromagnetic); biological effects of fields; biomedical measurement; calibration; measurement errors; measurement uncertainty; Antenessa; EME SPY 120; anechoic chamber; band-selective personal exposure meters; calibration measurement; measurement error; measurement uncertainty; Anechoic chambers; Calibration; Electromagnetic fields; Electromagnetic measurements; Frequency measurement; GSM; Linearity; Microwave measurements; Performance evaluation; Testing;
Conference_Titel :
Electromagnetic Compatibility (APEMC), 2010 Asia-Pacific Symposium on
Conference_Location :
Beijing
Print_ISBN :
978-1-4244-5621-5
DOI :
10.1109/APEMC.2010.5475751