DocumentCode :
2524355
Title :
Accuracy limits for simple molecular absorption based wavelength references
Author :
Swann, W.C. ; Gilbert, S.L.
Author_Institution :
National Institute of Standards and Technology, 325 Broadway, Boulder, CO 80305
fYear :
2004
fDate :
28-30 Sept. 2004
Firstpage :
15
Lastpage :
18
Abstract :
We present limitations on the characterization and use of molecular absorption lines for wavelength calibration purposes. These limits assume a Doppler and pressure-broadened absorption spectrum obtained by transmitting light through a simple gas cell. Attempting to define the line centers to ever-higher accuracy reveals limits associated with effects of cell pressure uncertainty, weak features in the spectrum, increasing complexity in the line profile model, and accuracy limits of wavelength meters.
Keywords :
Absorption; Calibration; Instruments; NIST; Optical sensors; Spectral analysis; Spectroscopy; Tunable circuits and devices; Uncertainty; Wavelength measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Optical Fiber Measurements, 2004. Technical Digest: Symposium on
Conference_Location :
Boulder, CO, USA
Print_ISBN :
1-886843-37-6
Type :
conf
DOI :
10.1109/SOFM.2004.183465
Filename :
1391918
Link To Document :
بازگشت