    A system for the realization of thermometric fixed points below 90K based on a closed-cycle He4 refrigerator is described. The goals of the cryostat design, low-temperature operation, instrumentation and data acquisition are explained. The current system is capable of achieving millikelvin (mK) accuracy in fixed-point determination. The significance of fixed-point realization in overall temperature calibration is explained
  • Keywords
    calibration; cryostats; liquid helium-4; low-temperature techniques; measurement standards; temperature measurement; 90 K; 4He; cryostat design; data acquisition; fixed-point determination; measurement standards; temperature calibration; thermometric fixed points; triple points; Calibration; Cryogenics; Data acquisition; Equations; Instruments; Interpolation; Platinum; Refrigeration; Temperature sensors; Thermodynamics;