Title :
Multilayer phased microcoil array for magnetic resonance imaging
Author :
Gruschke, O.G. ; Clad, L. ; Baxan, N. ; Kratt, K. ; Mohmmadzadeh, M. ; von Elverfeldt, D. ; Peter, A. ; Hennig, J. ; Badilita, V. ; Wallrabe, U. ; Korvink, J.G.
Author_Institution :
Dept. of Microsyst. Eng. - IMTEK, Univ. of Freiburg, Freiburg, Germany
Abstract :
We present for the first time wirebonded microcoils arranged in a planar phased-array configuration for large field of view (FOV) microscale magnetic resonance imaging (MRI) of 2D samples. The phased array consists of seven microcoils providing a sensitive area of 18.3 mm2. We demonstrate successful high-resolution imaging of a water phantom with 16 × 16 μm2 in-plane resolution and an signal-to-noise ratio (SNR) of 27 in 12 min 48 s acquisition time.
Keywords :
coils; lead bonding; magnetic resonance imaging; microfabrication; multilayers; 2D samples; MRI; high-resolution imaging; in-plane resolution; large field-of-view microscale; magnetic resonance imaging; multilayer phased microcoil array; planar phased-array configuration; signal-to-noise ratio; time 12.8 min; water phantom; wirebonded microcoils; Arrays; Coils; Gold; Magnetic resonance imaging; Voltage measurement; Water; Wires; MRI; Phased array; microcoil;
Conference_Titel :
Solid-State Sensors, Actuators and Microsystems Conference (TRANSDUCERS), 2011 16th International
Conference_Location :
Beijing
Print_ISBN :
978-1-4577-0157-3
DOI :
10.1109/TRANSDUCERS.2011.5969344