  • Abstract
    We report contraction-expansion array (CEA) microchannels for high-yield blood plasma separation utilizing differential inertial migration. An abrupt change of the cross-sectional area curves fluid streams and accelerates the flow velocity, which induces Dean flow. Utilizing the balance between inertial lift force and Dean drag force, we were able to separate blood plasma from human whole blood with a level of 62.2% yield. The CEA microfluidic device is easy to fabricate in a single layer fabrication process and expected to be useful for a simple blood plasma extraction on a chip with high-yield.
  • Keywords
    bioMEMS; biochemistry; blood; drag; haemodynamics; haemorheology; microchannel flow; microfabrication; CEA microfluidic device; Dean drag force; Dean flow; blood plasma extraction-on-a-chip; contraction-expansion array microchannel; cross-sectional area curves; flow velocity; fluid streams; high-yield blood plasma separation; inertial lift force; inertial migration; single layer fabrication process; Blood; Drag; Fluorescence; Force; Microchannel; Plasmas; Contraction-expansion array (CEA); Dean flow; Inertial microfluidics; Plasma separation;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Solid-State Sensors, Actuators and Microsystems Conference (TRANSDUCERS), 2011 16th International
  • Conference_Location
    Beijing
  • ISSN
    Pending
  • Print_ISBN
    978-1-4577-0157-3
  • Type

    conf

  • DOI
    10.1109/TRANSDUCERS.2011.5969423
  • Filename
    5969423