• DocumentCode
    2535484
  • Title

    Surgical instruments position detection system applied on minimally invasive insertion

  • Author

    David Gamino, C. ; Daniel Lorias, E. ; Arturo Minor, M.

  • Author_Institution
    Dept. de Electr., CINVESTAV-IPN, Mexico City, Mexico
  • fYear
    2012
  • fDate
    26-31 March 2012
  • Firstpage
    30
  • Lastpage
    34
  • Abstract
    In this paper, we show the results of a system that detects the surgical instruments applied to minimally invasive surgeries. This system is applied to the insertion of a 5 mm diameter needle inside a testing model; it obtains the 3D position of the needle, the lateral projection of the position and the frontal projection of the position. The system also determines the length of the instrument inside the model. The accuracy and precision of the system have been tested with results of 4% accuracy error and 3% precision error.
  • Keywords
    needles; position measurement; surgery; 3D position; accuracy error; frontal projection; lateral projection; minimally invasive insertion; needle; position detection system; precision error; surgical instruments; Conferences; Couplings; IEEE catalog; Instruments; Media; Surgery; 3D position; hybrid system; image processing; surgical technology;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Health Care Exchanges (PAHCE), 2012 Pan American
  • Conference_Location
    Miami, FL
  • Print_ISBN
    978-1-4673-1135-9
  • Type

    conf

  • DOI
    10.1109/PAHCE.2012.6233434
  • Filename