DocumentCode :
2537439
Title :
Analysis of the E-field Distribution for E-field Probes Calibration in Tissue-Simulating Liquid
Author :
Min, Yi ; Shufang, Li ; Hongjie, Liu ; Weijun, Hong
Author_Institution :
Beijing Univ. of Posts & Telecommun., Beijing
fYear :
2007
fDate :
23-26 Oct. 2007
Firstpage :
354
Lastpage :
357
Abstract :
In this paper, the E-field distribution in a waveguide partly filled with t issue-simulating liquid at mobile communications frequencies is presented. For the E-field probes calibration in the tissue-simulating liquid, a standing open rectangular waveguide partly filled with tissue-simulating liquid is used in which the E-field can be calculated analytically from input power measurements. It is desirable that the field distribution inside the tissue-simulating liquid follows the TE01 pattern. This has been verified by using the FDTD numerical technique and taking scan measurements at different heights from the dielectric spacer. The theoretical and simulating results are testified to be in good agreement with measurements.
Keywords :
biological effects of microwaves; calibration; finite difference time-domain analysis; probes; rectangular waveguides; E-field distribution; E-field probes calibration; FDTD numerical technique; dielectric spacer; finite different time domain technique; mobile communication; rectangular waveguide; tissue-simulating liquid; Calibration; Dielectric measurements; Frequency; Liquid waveguides; Mobile communication; Power measurement; Probes; Rectangular waveguides; Tellurium; Waveguide components; E-field Distribution; E-field Probe; Finite Different Time Domain; Rectangular Waveguide;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electromagnetic Compatibility, 2007. EMC 2007. International Symposium on
Conference_Location :
Qingdao
Print_ISBN :
978-1-4244-1371-3
Electronic_ISBN :
978-1-4244-1372-0
Type :
conf
DOI :
10.1109/ELMAGC.2007.4413503
Filename :
4413503
Link To Document :
بازگشت