Abstract :
Cochlear implants have become a standard treatment for many with severe to profound sensorineural hearing loss. However, delicate cochlear structures can be damaged during surgical insertion, which can lead to loss of residual hearing and decreased implant effectiveness. We propose a magnetic guidance concept in which a magnetically tipped cochlear implant is guided as it is inserted into the cochlea. In a scaled in vitro experimental study, we record insertion forces for nonguided and magnetically guided insertion experiments and compare the results. Results indicate that magnetic guidance reduced insertion forces by approximately 50%.
Keywords :
biology computing; ear; hearing; patient treatment; surgery; cochlear implants; delicate cochlear structures; magnetic guidance concept; sensorineural hearing loss; Force; Implants; Magnetic resonance imaging; Manipulators; Perpendicular magnetic recording; Prototypes; Torque;