Title :
Development of two dimensional thermal neutron flux monitor using multi-wire proportional counter for boron neutron capture therapy
Author :
Tanaka, Hiroya ; Sakurai, Yasushi ; Suzuki, M. ; Masunaga, Shin-ichiro ; Hino, Masami ; Kinashi, Yuko ; Ono, Keishi ; Maruhashi, Akira
Author_Institution :
Kyoto Univ. Res. Reactor Inst., Kumatori, Japan
fDate :
Oct. 27 2012-Nov. 3 2012
Abstract :
At Kyoto University Research Reactor Institute, clinical studies of around 400 of Boron Neutron Capture Therapy (BNCT) have been performed using research reactor. On the other hand, we have developed cyclotron based epithermal neutron source. To improve the quality of treatment, it is necessary to perform the detection of two dimensional thermal neutron distribution before and during treatment as quality control. Generally, the method of activation using gold foiVwire and germanium semiconductor detector is applied to detection of thermal neutron flux at several positions. However, the activation method using semiconductor detector takes some time to decide the thermal neutron flux. Therefore, we propose the method using multi-wire proportional counter for detection of two dimensional thermal neutron distribution. We constructed the system consisting of multi-wire detection head, multi-channel preamplifier, multi-channel discriminator, centroid calculation circuit, and data acquisition system. Detection efficiency for BNCT irradiation field was evaluated using Am-Be neutron source with moderator for producing thermal neutrons. Furthermore, two dimensional thermal neutron distribution was demonstrably obtained.
Keywords :
germanium radiation detectors; multiwire proportional chambers; neutron capture therapy; neutron sources; 2D thermal neutron distribution; 2D thermal neutron flux monitor; BNCT irradiation field; Kyoto University Research Reactor Institute; activation method; americium-beryllium neutron source; boron neutron capture therapy; centroid calculation circuit; cyclotron development; data acquisition system; epithermal neutron source; germanium semiconductor detector; multichannel discriminator; multichannel preamplifier; multiwire detection head; multiwire proportional counter; quality control; research reactor;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location :
Anaheim, CA
Print_ISBN :
978-1-4673-2028-3
DOI :
10.1109/NSSMIC.2012.6551089