DocumentCode :
2549263
Title :
Root Cause Analysis for medical equipment calibration laboratory nonconformities
Author :
Abdel Wahed, Manal ; Montaser, Mohamed ; Sami, Sherif A.
Author_Institution :
Syst. & Biomed. Eng. Dept. (SBMED), Cairo Univ., Cairo, Egypt
fYear :
2010
fDate :
16-18 Dec. 2010
Firstpage :
206
Lastpage :
209
Abstract :
Solving problems effectively is part of being an effective organization. Beneath every problem is a cause for that problem. In order to solve a problem one must identify the cause of the problem and take steps to eliminate the cause. Root Cause Analysis (RCA) is a method that is used to address a problem or non-conformance, in order to get to the “root cause” of the problem. One of the main requirements of medical equipment calibration laboratory accreditation process is the laboratory policy to address the problem of nonconformance of its activity. This paper presents an established and accredited procedure for the control of nonconforming calibration work and/or results. The procedure is based on the experience of the Medical Equipment Calibration Lab (MECL) of the Systems and Biomedical Engineering Department (SBME), Cairo University. A description of some nonconformity problem cases reports and their root cause analysis is also provided.
Keywords :
biomedical equipment; calibration; medical equipment calibration laboratory; nonconformities; root cause analysis; Biomedical engineering; Biomedical equipment; Calibration; IEC standards; ISO standards; Laboratories; Training;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Biomedical Engineering Conference (CIBEC), 2010 5th Cairo International
Conference_Location :
Cairo
ISSN :
2156-6097
Print_ISBN :
978-1-4244-7168-3
Type :
conf
DOI :
10.1109/CIBEC.2010.5716096