Abstract
The aim of this work is to provide the characteristics of MPPC arrays with an active area of 6x6 mm2 in gamma-ray spectrometry with 5x5x5mm3 CsI:TI scintillator at a wide temperature range. The measurements presented in this work covered: characteristics of breakdown voltage, selection of MPPC optimum operating voltage, verification of excess noise factor, measurement of energy resolution and the number of photoelectrons per 1 MeV. All measurements were carried out as a function of temperature and compared with similar results obtained previously with CsI:TI crystal coupled to a Large Area Photodiode (LAAPD).
Keywords
gamma-ray spectroscopy; photodiodes; solid scintillation detectors; CsI:TI crystal; CsI:TI scintillator; MPPC array; MPPC optimum operating voltage; breakdown voltage; energy resolution; excess noise factor; gamma-ray spectrometry; large area photodiode; operating voltage; photoelectrons; temperature -20 degC to 22 degC; Energy Resolution; Excess noise factor; MPPC; Non-proportionality; Number of photoelectron;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location
Anaheim, CA
ISSN
1082-3654
Print_ISBN
978-1-4673-2028-3
Type
conf
DOI
10.1109/NSSMIC.2012.6551430
Filename
6551430
Link To Document