Title :
The effect of receiving chamber environment on the visualization of cardiac jets by color Doppler flow mapping
Author :
Cape, Edward G. ; Monheit, Ruben ; Sung, Hsing-Wen ; Winoto, Sonny H. ; Weyman, Arthur E. ; Levine, Robert A. ; Yoganathan, Ajit P.
Author_Institution :
Sch. of Chem. Eng., Georgia Polytech, Atlanta, GA, USA
Abstract :
Jet spatial resolution as visualized by color Doppler flow mapping has previously been shown to be affected by physical factors, such as driving pressure, and technical factors, such as instrument settings. An additional factor producing superimposed variability, namely, the interaction of the jet with solid boundaries, in the distal receiving chamber, is illustrated. The broader point that jet velocities (i.e., Doppler images) do not uniquely represent jet volume is demonstrated
Keywords :
Doppler effect; biomedical ultrasonics; cardiology; haemodynamics; jets; cardiac jets visualization; color Doppler flow mapping; distal receiving chamber; driving pressure; instrument settings; physical factors; receiving chamber environment; solid boundaries; technical factors; Blood flow; Cells (biology); Frequency; Heart; Instruments; Solids; Ultrasonic imaging; Valves; Velocity measurement; Visualization;
Conference_Titel :
Visualization in Biomedical Computing, 1990., Proceedings of the First Conference on
Conference_Location :
Atlanta, GA
Print_ISBN :
0-8186-2039-0
DOI :
10.1109/VBC.1990.109318