DocumentCode :
2554118
Title :
Imaging performance of DOI measurable PET systems for breast imaging: Monte Carlo simulation studies
Author :
Jing Wu ; Xishan Sun ; Kai Lou ; Yan Xia ; Tianyu Ma ; Yiping Shao
Author_Institution :
Dept. of Eng. Phys., Tsinghua Univ., Beijing, China
fYear :
2012
fDate :
Oct. 27 2012-Nov. 3 2012
Firstpage :
2274
Lastpage :
2277
Abstract :
GATE based simulation studies were conducted to guide breast PET system designs with different system geometries. A breast phantom (modeled as a half prolate spheroid with 7 cm equatorial radii and 11 cm polar radius) was generated for the study with warm background and hot lesions of different diameters (2, 3, 4 and 6 mm) and lesion/background activity ratios (3:1, 5:1 and 8:1, F-18 isotope). Each detector consists of an 8 × 8 array of 2 mm × 2 mm × 30 mm LSO scintillators with depth-of-interaction (DOI) measurement capability. Four systems were simulated: two stationary systems both with four detector panels arranged in a box-shaped geometry but different panel-to-panel separations of 17.2 cm and 20.4 cm, a system with two rotating detector panels separated at 17.2 cm panel-to-panel distance, and a whole-body (WB) PET with ~90 cm detector ring diameter. Axial FOVs are 12.8 cm for all three breast systems and 16 cm for WB PET. A list-mode OSEM algorithm with 40 subsets and one-pass iteration was used for image reconstruction. The reconstructed images and measured contrast-noise-ratio (CNR) with different system configurations, DOI resolutions, lesion size, and activity ratios were used to compare the imaging performance of different systems. The results validate that DOI is critical to providing uniform spatial resolution across the FOV for a breast PET system with compact geometry, and show that the full-tomographic configuration with detectors encompassing the FOV provides superior imaging performance than that from the rotating detector PET or WB PET, with very high sensitivity (> 60% at the center of FOV), much better lesion visual identification, and significantly improved CNR. Based on these quantitative evaluations, this simulation study has demonstrated that even coarse DOI resolution (~5 mm) can provide substantially improved imaging performance for a small bore system and design flexbility with different system geometries.
Keywords :
Monte Carlo methods; geometrical optics; image reconstruction; positron emission tomography; scintillation counters; DOI measurable PET system; DOI resolution; GATE based simulation; LSO scintillators; Lu2SiO5:Ce; Monte Carlo simulation; axial FOV; box-shaped geometry; breast PET system design; breast imaging; breast phantom; compact geometry; contrast-noise-ratio; depth-of-interaction measurement; detector ring diameter; four detector panels; full-tomographic configuration; image reconstruction; imaging performance; lesion size; lesion/background activity ratios; list-mode OSEM algorithm; one-pass iteration; panel-to-panel distance; panel-to-panel separation; rotating detector panels; small bore system; whole-body PET;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location :
Anaheim, CA
ISSN :
1082-3654
Print_ISBN :
978-1-4673-2028-3
Type :
conf
DOI :
10.1109/NSSMIC.2012.6551516
Filename :
6551516
Link To Document :
بازگشت