Title :
Use of scattered coincidences for emission-based estimation of attenuation map in PET
Author :
Berker, Yannick ; Kiessling, F. ; Schulz, V.
Author_Institution :
Univ. Hosp., Dept. of Exp. Mol. Imaging, RWTH Aachen Univ., Aachen, Germany
fDate :
Oct. 27 2012-Nov. 3 2012
Abstract :
Attenuation correction in PET/MR is difficult due to the lack of X-ray or radionuclide transmission scans. Known approaches to estimating attenuation maps from PET emission data are based on true coincidences only and fail in highly symmetric situations. We present a method of extracting information about the attenuation distribution from scattered coincidences, even with perfect spherical symmetry of attenuation and activity. Using GATE simulations in a realistic scanner geometry combined with perfect energy resolution, we show how back-projections of scattered coincidences can provide an estimate of the attenuation map. This method may ultimately be used to complement existing approaches, such as for compensation of artifacts in segmentation-based attenuation correction, or for determination of a missing constant in Maximum Likelihood reconstruction of Activity and Attenuation.
Keywords :
medical signal processing; positron emission tomography; scattering; signal reconstruction; GATE simulation; PET attenuation map; PET emission data; attenuation correction; attenuation distribution; emission based estimation; perfect energy resolution; perfect spherical symmetry; photon attenuation; realistic scanner geometry; scattered coincidence back projection;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location :
Anaheim, CA
Print_ISBN :
978-1-4673-2028-3
DOI :
10.1109/NSSMIC.2012.6551519