Title :
Comparative transient and steady-state performances of three resonant PWM inverter-controlled DC-DC converters using high-voltage transformer and cable parasitic circuit components
Author :
Sun, J.M. ; Gamage, L. ; Nakaoka, M. ; Takano, H. ; Hatakeyama, T.
Author_Institution :
Graduate Sch. of Eng. & Sci., Yamaguchi Univ., Japan
Abstract :
The main purpose of this study is to describe some particular characteristic features and inherent benefits of series, parallel and series-parallel resonant tank inverter-fed DC-DC converters for a medical-use X-ray high-voltage generator from a practical viewpoint. The method to treat the parasitic resonant circuit parameters of the high-voltage high-frequency transformer used in three typical resonant tank topologies is discussed on the basis of some simulation results, and the simulation results are also verified by experiments
Keywords :
DC-DC power convertors; PWM power convertors; X-ray production; X-ray tubes; biomedical equipment; diagnostic radiography; invertors; power supplies to apparatus; resonant power convertors; transformers; cable parasitic circuit components; high-voltage high-frequency transformer; high-voltage transformer; medical-use X-ray high-voltage generator; parallel converter; resonant PWM inverter-controlled DC-DC converters; resonant tank inverter-fed DC-DC converters; series converter; series-parallel converter; steady-state performance; transient performance; DC-DC power converters; Oil insulation; Parasitic capacitance; Power generation; Pulse width modulation inverters; RLC circuits; Resonance; Resonant inverters; Steady-state; X-ray imaging;
Conference_Titel :
Power Conversion Conference - Nagaoka 1997., Proceedings of the
Conference_Location :
Nagaoka
Print_ISBN :
0-7803-3823-5
DOI :
10.1109/PCCON.1997.638336