• DocumentCode
    2554383
  • Title

    Two algorithms for correcting phase aberration in a computer-controlled ultrasonic imaging system

  • Author

    Zhao, Danhua ; Trahey, Gregg E.

  • Author_Institution
    Dept. of Biomed. Eng., Duke Univ., Durham, NC, USA
  • fYear
    1990
  • fDate
    3-6 Jun 1990
  • Firstpage
    38
  • Lastpage
    43
  • Abstract
    Spatial inhomogeneities in the acoustic velocity of tissues degrade the performance of medical ultrasonic imaging systems. Two algorithms which can correct the phase aberration induced by the inhomogeneities are presented. The first algorithm adjusts the phases of signals transmitted and received from individual channels in a phased array scanner, one group of adjacent elements at a time. The second algorithm is based on the assumption that a phases error profile can be approximated by a sum of sine waves and the phase aberration can be corrected by subtracting components of the sum. Comparisons of the two algorithms are presented
  • Keywords
    acoustic imaging; biomedical ultrasonics; computerised instrumentation; computerised picture processing; medical diagnostic computing; computer-controlled ultrasonic imaging system; medical ultrasonic imaging systems; phase aberration correction algorithms; phases error profile; sine waves sum; spatial inhomogeneities; tissue acoustic velocity; Biomedical acoustics; Biomedical computing; Brightness; Degradation; Error correction; Frequency; Optical imaging; Phased arrays; Ultrasonic imaging; Ultrasonic transducer arrays;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computer-Based Medical Systems, 1990., Proceedings of Third Annual IEEE Symposium on
  • Conference_Location
    Chapel Hill, NC
  • Print_ISBN
    0-8186-9040-2
  • Type

    conf

  • DOI
    10.1109/CBMSYS.1990.109376
  • Filename
    109376