• DocumentCode
    2555956
  • Title

    Stationary SPECT with multi-layer Multiple-Pinhole-Arrays

  • Author

    Ren, Weiqiang ; Valastyan, I. ; Colarieti-Tosti, M.

  • Author_Institution
    Dept. of Med. Eng., R. Inst. of Technol. (KTH), Huddinge, Sweden
  • fYear
    2012
  • fDate
    Oct. 27 2012-Nov. 3 2012
  • Firstpage
    2594
  • Lastpage
    2597
  • Abstract
    The potential of Multiple Pinhole Arrays (MPA) collimators for developing a Single Photon Emission Computer Tomography (SPECT) system without rotating or moving elements is investigated. A four layer arrangement is proposed and the system performance is evaluated using the simulation toolkit GATE [1]. For a camera with a field of view (FOV) of the order of a human brain (a sphere of radius 100 mm), a sensitivity of 86,0 cps/MBq and an overall resolution of 5 mm have been estimated, indicating that performances comparable to traditional parallel-hole-collimator cameras can be achieved.
  • Keywords
    brain; cameras; collimators; single photon emission computed tomography; FOV; Single Photon Emission Computer Tomography system; camera; field-of-view; human brain; multilayer multiple-pinhole-array; parallel-hole-collimator cameras; simulation toolkit GATE; stationary SPECT;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
  • Conference_Location
    Anaheim, CA
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-2028-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2012.6551593
  • Filename
    6551593